Advertisement

Continuous renal replacement therapy: forty-year anniversary

Abstract

In 1977 Peter Kramer performed the first CAVH (continuous arteriovenous hemofiltration) treatment in Gottingen, Germany. CAVH soon became a reliable alternative to hemo- or peritoneal dialysis in critically ill patients. The limitations of CAVH spurred new research and the discovery of new treatments such as CVVH and CVVHD (continuous veno-venous hemofiltration and continuous veno-venous hemodialysis). The alliance with industry led to development of new specialized equipment with improved accuracy and performance in delivering continuous renal replacement therapies (CRRTs). Machines and filters have progressively undergone a series of technological steps, reaching a high level of sophistication. The evolution of technology has continued, leading to the development and clinical application of new membranes, new techniques and new treatment modalities. With the progress of technology, the entire field of critical care nephrology moved forward, expanding the areas of application of extracorporeal therapies to cardiac, liver and pulmonary support. A great deal of research made extracorporeal therapies an interesting option for the treatment of sepsis and intoxication and the additional use of sorbents was explored. With the progress in understanding the pathophysiology of acute kidney injury (AKI), new guidelines were developed, driving indications, modalities of prescription, monitoring techniques and quality assurance programs. Information technology and precision medicine have recently contributed to further evolution of CRRT, with the possibility of collecting data in large databases and evaluating policies and practice patterns. This is likely to ultimately result in improved patient care. CRRTs are 40 years old today, but they are still young and full of potential for further evolution.

Int J Artif Organs 2017; 40(6): 257 - 264

Article Type: REVIEW

DOI:10.5301/ijao.5000610

OPEN ACCESS ARTICLE

Authors

Claudio Ronco

Article History

Disclosures

Financial support: None.
Conflict of interest: None.

This article is available as full text PDF.

Download any of the following attachments:

Introduction

Exactly forty years ago Peter Kramer treated the first patient with continuous arterio-venous hemofiltration (CAVH) in the intensive care unit (ICU) of Gottingen, Germany (1). Acute renal failure was mostly treated with peritoneal dialysis or haemodialysis, but in critically ill patients, these modalities were often contraindicated or precluded due to severe cardiovascular instability. CAVH was well tolerated and easy to institute in ICUs where haemodialysis was not routinely performed. In the year following the first description of CAVH, a significant evolution of acute renal replacement therapy (ARRT) and related technology occurred, leading to the modern techniques of continuous renal replacement therapy (CRRT). In this process of evolution, the centre of Vicenza has always been leading the edge and it has significantly contributed to the birth of a new speciality called Critical Care Nephrology. In Figure 1 we report the history timeline of CRRT as seen from our point of observation. It may be incomplete or partial, but certainly every single minute of this history has been personally lived with passion and dedication.

Timetable of the forty years of evolution in CRRT.

The era of CAVH

The introduction of CAVH made it possible to perform renal replacement therapy even in ICUs not fully equipped or trained for hemodialysis. Our first patient was treated in Vicenza with the original Amicon 20 diafilter in 1978. We further refined our understanding and knowledge of CAVH by collaborating with Juan Bosch at the Mount Sinai Hospital in New York (2). In the late 1970s we extended the use of the technique used in adults to neonates with the use of specific Minifilters (Fig. 2) (3, 4). We developed specific filters with reduced flow resistance, suitable for operating in arterio-venous mode that are today preserved in our Vicenza Dialysis Museum. In spite of several advantages, CAVH soon demonstrated limitations in efficiency and frequent complications due the need for arterial cannulation. This spurred new research for alternative techniques.

Vicenza, 1978: First CAVH treatment in an adult critically ill patient and first CAVH treatment in the world of a neonate with acute renal failure.

The discovery of CAVHD and CAVHDF

A remarkable increase in treatment efficiency and urea removal was achieved with the addition of diffusion. New filters with 2 ports in the dialysate/filtrate compartment allowed the use of counter-current flow of dialysate, giving birth to continuous arteriovenous hemodiafiltration or hemodialysis (CAVHDF or CAVHD) (5). CAVHD- CAVHDF made it possible to treat hyper-catabolic patients, simply by increasing dialysate the flow rate up to 1.5 or 2 L/h. In the case of excessive ultrafiltration, fluid losses were partially or completely replaced, allowing accurate fluid balance control. The problem of arterial cannulation was still the main drawback of the technique.

Fluid balance control

Manual control of ultrafiltration was initially achieved by positioning the filtrate bag at different heights, thus modifying the negative (suction) pressure generated by the filtrate column. Delivery of replacement fluid was initially regulated manually and later new systems were designed to provide automatic fluid balance (6). The systems operated by gravity, using scales and electronic clamps, although peristaltic pumps soon replaced these simple mechanisms with more advanced equipment.

CVVH, CVVHD and CVVHDF

Because arterial-venous techniques were associated with significant complications related to arterial cannulation and low blood flow in the extracorporeal circulation, continuous veno-venous hemofiltration (CVVH) or hemodialysis (CVVHD) hemodiafiltration (CVVHDF) took over thanks to the availability of double lumen venous catheters and machines with a peristaltic blood pump (7-8-9-10). These changes improved the safety and performance of extracorporeal treatments in the ICU, allowing the use of high blood flows (>150 mL/min) and large dialyzers (>1.0 m2). This resulted in better control of hyper-catabolic states and other severe pathophysiological derangements. Initially veno-venous circuits took advantage of already existing technology, often creating a combination of devices (blood pump, UF pump, reinfusion pump, anticoagulation, etc.) that were not integrated and, therefore, unable to communicate and operate together, with the possibility of amplifying errors and complications. Adoptive technology allowed significant advances to be made, but it soon proved inadequate when a clear need for integrated and safer devices became evident in clinical practice.

Integrated technology

In Vicenza we started the process of integration of blood and dialysate filtrate pumps in a single machine and pushed the industry of the field toward new research to create dedicated devices for CRRT. With this effort, companies mostly involved in chronic hemodialysis began to collect components from chronic machines and to integrate them into machines for CRRT (11, 12). Using this approach, several companies made new equipment: B.Braun moved from TRIO (3 separate pumps) to the compact ECU Carex machine (Fig. 1) designed by the engineer, Mr. Frigato, in Mirandola. Baxter moved to the integrated BM 25; and Hospal generated the DM 32 from the blood module of the Monitral machine. Medica built an integrated version of the Equapump and Equaline, while in Germany, Fresenius Medical Care commercializsed the DM08 with 1 blood pump and 2 pumps for dialysate or fluid balance (13-14-15). Bellco utilized the blood module of the Multimat chronic machine to create the Multimat B Acute version, and Gambro adjusted a special version of the AK 10 module for continuous dialysis. All these machines derived from the technology generated for chronic hemodialysis, but they represented the seeds for the upcoming generation of specific CRRT equipment (16, 17).

The PRISMA® revolution

With the progress in understanding the pathophysiology of AKI and its clinical implications, the targets for renal replacement therapy became clearer (18-19-20-21-22-23-24-25). The indications for CRRT were defined and the practice of CRRT was better identified in the first edition of the book Critical Care Nephrology, edited by Bellomo and Ronco. New standard of treatments were required as well as new simplified methods to expand the application of CRRT to all ICUs. Special requirements for easier institution of CRRT and easier monitoring of treatment had led to the development of the first generation of CRRT integrated machines with several pumps and different technique capabilities. It was now the time for a revolution with a second-generation CRRT machine, and this happened with the launch of the PRISMA®, the first integrated CRRT platform designed specifically for Acute RRT in intensive care. The pre-assembled circuit and the auto-priming feature, together with the 4 pumps made CRRT possible in almost every ICU, with improved safety and performance even when performing complex techniques such as CVVHDF.

CRRT dose quantification

Started in 1995, in Vicenza we concluded the first randomized trial on treatment dose in the year 2000 (26). We underlined the importance of a minimum quantity (dose) of treatment and we indicated 35 mL/kg per hour as the optimal target according to the trial result. This target, with a minimum of 80% delivery of the prescribed dose, began the golden standard for adequate CRRT and improved clinical outcomes. Subsequent studies have demonstrated that lower doses can be equally safe and successful in treating the critically ill patient (27-28-29-30-31-32-33), although effective delivery often differs significantly from the prescription (34-35-36-37). The new machines developed by industry with the possibility of moving toward more sophisticated techniques such as coupled plasma filtration-adsorption (CPFA) and high-volume hemofiltration (HVHF) fulfilled these new performance requirements (38-39-40-41). Improved usability and user-friendly interfaces with systematic instructions made CRRT machines popular and widely used in every ICU (17).

CRRT in sepsis

The original observation that higher doses of treatment could be beneficial in patients with sepsis introduced the rationale for HVHF and CPFA with the intent to remove chemical mediators (42-43-44-45). HVHF or CPFA showed potent immunomodulatory effects in sepsis. Since a cytokine network that is synergistic, redundant, autocatalytic and self-augmenting characterizes sepsis and systemic inflammatory response syndrome, it is unlikely to achieve the control of such a nonlinear system by simple blockade or elimination of some specific mediators. Thus, we hypothesized that the unselective nature of mediator removal by extracorporeal therapies could be a potential advantage, capable of reconstituting the immune-homeostasis of the septic patient, as suggested by the ‘peak concentration’ hypothesis (42). Higher volumes and complex techniques were made possible by the third generation of CRRT machines such as the Prismaflex (Gambro), the Equasmart (Medica), The Lynda (Bellco), The Multifiltrate (Fresenius Medical Care), The Acquarius (Edwards Life Sciences) and others.

From renal replacement to multiple organ support therapy

The effect of different modalities of CRRT on length of stay and recovery of renal function in the general population is still under evaluation, since the case mix changes in every study and the population treated is not homogeneous. Further research is needed in this field, although it has become evident that Precision CRRT (personalized prescription) should be adopted in order to optimize results in single patients, even in the absence of the documented benefits of 1 specific technology for the general population. Adequate technological support becomes mandatory to fulfil all these expectations and new machines have been upgraded with specific circuits to support organs other than the kidney. The possibility of supporting or partially replacing the function of organs such as the liver, the lung and the heart with modified extracorporeal circuits, make multiple organ support therapy (MOST) a reality in the critical care setting. CRRT machines become platforms for multiple organ support via extracorporeal therapies carried out with specific biomaterials and devices (45, 46).

Sorbent therapies

Hemoperfusion has been available for detoxification purposes since the second part of the last century. Zeolites and charcoal were the most common sorbents, although soon a new generation of more hemocompatible materials became available. In recent years, cartridges with polymixin-B-coated polystyrenic fibres designed for the adsorption of endotoxin have been used for the treatment of sepsis. This therapy seems particularly beneficial in patients with post-surgery abdominal septic shock (47, 48). New sorbents, derived from the original experience carried out by our group in the early years of the new century, are today appearing as a new option for the treatment of sepsis (49-50-51-52-53).

Latest generation of CRRT machines

The latest generation of machines available on the market today, which represent the evolution of the past 4 decades of research and development is shown in Figure 3. Specific machines have now been designed to permit safe and reliable performance of the therapy. These new devices are equipped with a user-friendly interface that allows for easy performance and monitoring. The apparent complexity of the circuit is made simple by a self-loading circuit or a cartridge that includes the filter and the blood and dialysate lines. Priming is performed automatically by the machine and pre- or post-dilution (reinfusion of substitution fluid before or after the filter) can easily be performed by changing the position of the reinfusion line. These new machines permit all CRRT techniques to be performed by programming the flows and the total amounts of fluid to be exchanged or circulated as a counter-current dialysate throughout the session.

Timetable of the forty years of evolution in pediatric CRRT.

Precision CRRT

The Acute Disease Quality Initiative consensus group has stressed the importance of applying the principles of precision medicine to CRRT (54). The group has invited CRRT experts and manufacturers to utilize a standardized nomenclature and sign a document called “Charta of Vicenza” (55) agreeing to follow the recommendations proposed by 2 detailed publications (56, 57). This was a starting point to generate a harmonized terminology and to advance in a standardized environment for the generation of useful pragmatic trials. Today we have several examples of new machines that include software and terminology coherent with the recommended nomenclature.

The consensus group also focused on the identification of patients requiring CRRT and the correct timing of application, technological needs and expected advances, pointing out the desirable characteristics of new equipment, membranes and the importance of integration of information technology into the process of patient care and decision-making (58-59-60-61).

Information technology and connectivity

The ADQI group has recently recommended the use of modern information technology (IT) tools to improve practice and patient care. Such tools can be used for data acquisition and storage but also for treatment monitoring (62). Manual, authorized or automatic feedback technology is available today in chronic dialysis machines and has been advocated for CRRT machines as well. Integration of patient and machine signals through IT tools and connectivity with electronic health record (EMR) and data collection systems will be required to allow pragmatic trials and make big data registries available for analysis. Data can then be used for Quality Assurance (QA) and for Continuous Quality Improvement (CQI) purposes in the centre, in the region and even in multinational data collection studies.

Machine connectivity can be provided via different tools. Machine and patient chip-cards can be used to extract data from single treatments from the front-end terminal (CRRT machine). Cable or wireless connectivity may permit the download of technical and clinical parameters from single or multiple machines and to analyse single treatment data as well as trends or statistics in multiple treatments. Cloud-based connectivity could help clinicians to generate virtual registries and analyse the performance of single treatments or centres in absolute terms, or relative to other units. This may result in important feedback to clinicians, to either strictly control outliers, or to change policies and procedures in the event of repeatedly unsatisfactory results.

Data collected and stored in EMRs may be rapidly evaluated and managed by ad hoc designed electronic sniffers that can alert clinicians about dangerous trends or unwanted effects of CRRT. Solutions to the problem may be listed as suggestions or even automatically fed back into devices such as pumps and CRRT machines. The feedback may either require manual application of the necessary change by a nurse/physician, authorization of an operational change proposed by the system, or operate automatically.

Current and future perspectives

There are several areas of investigation where evolution in new therapeutic options and new devices is likely to occur. Different biomaterials and surface modifications have been introduced on semipermeable membranes to increase biocompatibility, to reduce thrombogenicity and the tendency toward fouling, and finally to modify sieving and adsorption properties. Vitamin E-coated membranes represent an example in which alpha-tocopherol has been covalently bound to a polysulfone membrane to reduce oxidant stress (63). These and other new membranes should be further investigated to elucidate the potential for improving the inflammatory pattern in sepsis.

Software integration should become a tool for new equipment to perform safer and more efficient CRRT treatments. The integration of bioimpedance and on-line haematocrit measurements may result in important effects on hemodynamic stability and treatment delivery (64). The integration of blood temperature monitoring may have the potential to control thermal energy balance and the patient’s temperature. This could allow a specific energy balance (KJ/h) or a target temperature control to be achieved by adjusting the dialysate or replacement fluid temperature according to signals coming from temperature sensors placed on blood and dialysate lines. A significant heat loss can in fact occur when the extracorporeal circuit is exposed to room temperature (65).

Automated circuit pressure control with flow adjustment feedback could provide the best possible blood flow in presence of a malfunctioning catheter or could provide early warning of access malfunction, preventing inadequate treatment delivery.

On-line chemical sensors for acid-base and electrolytes may provide the basis for continuous control via biochemical feedback on dialysate and replacement fluid composition.

Miniaturized and wearable technology

There is a great deal of interest in applying nanotechnology, microfluidics and other emerging sciences to the field of renal replacement. In Vicenza we tested the first wearable system for ultrafiltration (66). For the original wearable belt, we mostly used components off the shelf, but we recently made new developments by creating a new design and applying newly conceived, dedicated technology (67-68-69). Wearable devices are mostly conceived for ambulatory care and out-of-hospital patients. Nevertheless, it seems that miniaturized systems may become useful even in the acute patient, allowing mobility, low-priming-volume extracorporeal circuits and better care at the bedside.

Specific technology for small infants and paediatric applications

An interesting development in the process of machine and device miniaturization is the application of small-sized components for infants and children for whom the adult technology seems to be clearly inadequate. The requirement for specific technology in the field of paediatric CRRT emerged in the early 1980s (70). Over the years we personally contributed to a significant evolution of the field (71-72-73-74) and the historical milestones of paediatric CRRT are reported in Figure 3. Quite recently, based on the observation that every CRRT machine is conceived for adult patients and the use in neonates and small infants is precluded or off-label, we decided to undertake a special project for the creation of miniaturized equipment specifically designed for these small patients. We started from scratch and we moved, through artisanal design and rough assembly of prototypes, to an industrially designed machine called CARPEDIEM (Cardio Renal Pediatric Dialysis Emergency Machine). This evolution has captured new interest in the treatment of AKI in neonates and has allowed different techniques to be performed with the possibility of precisely adjusting the treatment dose and fluid balance in babies with body weights as low as 1.5 kg (75, 76) (Fig. 4). There is a consensus that such a specific technology will probably modify the outcome of AKI in neonates and small infants.

First CRRT treatment with CARPEDIEM (Cardio renal Pediatric Dialysis Emergency Machine).

Conclusions

Over the last 40 years, CRRT has been widely utilized for the management of AKI in critically ill patients (77-78-79). The Vicenza Centre has heavily contributed to important technological advances in machine design, to the expanded application of novel and complex modalities and, finally, to the optimization of extracorporeal therapies in the ICU, ultimately leading to improved patient outcomes.

We pushed the transition from CRRT as renal replacement therapy to the concept of MOST (Multiple Organ Support Therapy) (80), upgrading the extracorporeal circuit to provide support for patients with multiple organ dysfunctions. Information and communication technology together with big data collection represent the new frontier (81-82-83-84-85) for personalized prescriptions and measurement of results to achieve the desired level of precision CRRT (86).

I strongly recommend using the history of the last 40 years of CRRT not only for a simple celebration of an important anniversary, but also for building the future of our discipline with improved treatment of acute kidney injury and organ dysfunction in the critically ill patient.

Disclosures

Financial support: None.
Conflict of interest: None.
References
  • 1. Kramer P Schrader J Bohnsack W Grieben G Gröne HJ Scheler F Continuous arteriovenous haemofiltration. A new kidney replacement therapy. Proc Eur Dial Transplant Assoc 1981 18 743 749 Google Scholar
  • 2. Lauer A Saccaggi A Ronco C Belledonne M Glabman S Bosch JP Continuous arterio-venous hemofiltration in the critically ill patient: Clinical Use and Operational Characteristics. Ann Intern Med 1983 99 455 460 Google Scholar
  • 3. Ricci Z Ronco C Technical advances in renal replacement therapy. Semin Dial 2011 24 2 138 141 Google Scholar
  • 4. Ronco C Garzotto F Ricci Z CA.R.PE.DI.E.M. (Cardio-Renal Pediatric Dialysis Emergency Machine): evolution of continuous renal replacement therapies in infants. A personal journey. Pediatr Nephrol 2012 27 8 1203 1211 Google Scholar
  • 5. Ronco C Arterio-venous hemodiafiltration (A-V HDF): a possible way to increase urea removal during C.A.V.H. Int J Artif Organs 1985 8 1 61 62 Google Scholar
  • 6. Ronco C Fabris A Feriani M et al. Technical and clinical evaluation of a new system for ultrafiltration control during hemodialysis. ASAIO Trans 1988 34 3 613 616 Google Scholar
  • 7. Zobel G Ring E Kuttnig M Grubbauer HM Five years experience with continuous extracorporeal renal support in paediatric intensive care. Intensive Care Med 1991 17 6 315 319 Google Scholar
  • 8. Better OS Acute renal failure in casualties of mass disasters. Kidney Int Suppl 1993 41 S235 S236 Google Scholar
  • 9. Werner HA Herbertson MJ Seear MD Operating characteristics of pediatric continuous arteriovenous hemofiltration in an animal model. Pediatr Nephrol 1993 7 2 189 193 Google Scholar
  • 10. Heinrichs W Mönk S Fauth U Halmágyi M An automatic system for fluid balance in continuous hemofiltration with very high precision. Contrib Nephrol 1991 93 167 170 Google Scholar
  • 11. Akcahuseyin E Vincent HH van Ittersum FJ van Duyl WA Schalekamp MA A mathematical model of continuous arterio-venous hemodiafiltration (CAVHD). Comput Methods Programs Biomed 1990 31 3-4 215 224 Google Scholar
  • 12. Barber ND Gibbs JS Barrett P Fox KM Quality of haemofiltration fluids: a potential cause of severe electrolyte imbalance. Br Med J (Clin Res Ed) 1987 295 6605 1025 Google Scholar
  • 13. Colton CK Analysis of membrane processes for blood purification. Blood Purif 1987 5 4 202 51 Google Scholar
  • 14. Macias WL Mueller BA Scarim SK Robinson M Rudy DW Continuous venovenous hemofiltration: an alternative to continuous arteriovenous hemofiltration and hemodiafiltration in acute renal failure. Am J Kidney Dis 1991 18 4 451 458 Google Scholar
  • 15. Yorgin PD Krensky AM Tune BM Continuous venovenous hemofiltration. Pediatr Nephrol 1990 4 6 640 642 Google Scholar
  • 16. Ricci Z Bonello M Salvatori G et al. Continuous renal replacement technology: from adaptive devices to flexible multipurpose machines. Crit Care Resusc 2004 6 3 180 187 Google Scholar
  • 17. Ronco C Bellomo R The evolving technology for continuous renal replacement therapy from current standards to high volume hemofiltration. Curr Opin Crit Care 1997 3 426 433 Google Scholar
  • 18. Tan HK Baldwin I Bellomo R Continuous veno-venous hemofiltration without anticoagulation in high-risk patients. Intensive Care Med 2000 26 11 1652 1657 Google Scholar
  • 19. Abramson S Niles JL Anticoagulation in continuous renal replacement therapy. Curr Opin Nephrol Hypertens 1999 8 6 701 707 Google Scholar
  • 20. Böhler J Donauer J Keller F Pharmacokinetic principles during continuous renal replacement therapy: drugs and dosage. Kidney Int Suppl 1999 72 S24 8 Google Scholar
  • 21. Jacka MJ Ivancinova X Gibney RT Continuous renal replacement therapy improves renal recovery from acute renal failure. Can J Anaesth 2005 52 3 327 332 Google Scholar
  • 22. Gillespie RS Seidel K Symons JM Effect of fluid overload and dose of replacement fluid on survival in hemofiltration. Pediatr Nephrol 2004 19 12 1394 1399 Google Scholar
  • 23. Luyckx VA Bonventre JV Dose of dialysis in acute renal failure. Semin Dial 2004 17 1 30 36 Google Scholar
  • 24. Clark WR Turk JE Kraus MA Gao D Dose determinants in continuous renal replacement therapy. Artif Organs 2003 27 9 815 820 Google Scholar
  • 25. Liao Z Zhang W Hardy PA et al. Kinetic comparison of different acute dialysis therapies. Artif Organs 2003 27 9 802 807 Google Scholar
  • 26. Ronco C Bellomo R Homel P et al. Effects of different doses in continuous venovenous haemofiltration on outcomes of acute renal failure: a prospective randomised trial. Lancet 2000 356 26 30 Google Scholar
  • 27. Ricci Z Salvatori G Bonello M et al. A new machine for continuous renal replacement therapy: from development to clinical testing. Expert Rev Med Devices 2005 2 1 47 55 Google Scholar
  • 28. Garzotto F Ostermann M Martín-Langerwerf D et al. DoReMIFA study group. The Dose Response Multicentre Investigation on Fluid Assessment (DoReMIFA) in critically ill patients. Crit Care 2016 20 1 196 Google Scholar
  • 29. Lorenzin A Garzotto F Alghisi A et al. CVVHD treatment with CARPEDIEM: small solute clearance at different blood and dialysate flows with three different surface area filter configurations. Pediatr Nephrol 2016 31 10 1659 1665 Google Scholar
  • 30. Villa G Ricci Z Romagnoli S Ronco C Multidimensional Approach to Adequacy of Renal Replacement Therapy in Acute Kidney Injury. Contrib Nephrol 2016 187 94 105 Google Scholar
  • 31. Swartz R Pasko D O’Toole J Starmann B Improving the delivery of continuous renal replacement therapy using regional citrate anticoagulation. Clin Nephrol 2004 61 2 134 143 Google Scholar
  • 32. Ricci Z Ronco C Renal replacement therapy in the critically ill: getting it right. Curr Opin Crit Care 2012 18 6 607 612 Google Scholar
  • 33. Ricci Z Ronco C Timing, dose and mode of dialysis in acute kidney injury. Curr Opin Crit Care 2011 17 6 556 561 Google Scholar
  • 34. Vesconi S Cruz DN Fumagalli R et al. DOse REsponse Multicentre International collaborative Initiative (DO-RE-MI Study Group). Delivered dose of renal replacement therapy and mortality in critically ill patients with acute kidney injury. Crit Care 2009 13 2 R57. Google Scholar
  • 35. Ronco C Cruz D Oudemans van Straaten H et al. Dialysis dose in acute kidney injury: no time for therapeutic nihilism - a critical appraisal of the Acute Renal Failure Trial Network study. Crit Care 2008 12 5 308 Google Scholar
  • 36. Ricci Z Bellomo R Ronco C Dose of dialysis in acute renal failure. Clin J Am Soc Nephrol 2006 1 3 380 388 Google Scholar
  • 37. Monti G Herrera M Kindgen-Milles D et al. Dose Response Multicentre International Collaborative Initiative Scientific Committee. The DOse Response Multicentre International Collaborative Initiative (DO-RE-MI). Contrib Nephrol 2007 156 434 443 Google Scholar
  • 38. Reiter K Bellomo R Ronco C Kellum J Pro/con clinical debate: Is high-volume hemofiltration beneficial in the treatment of septic shock? Crit Care 2002 6 18 21 Google Scholar
  • 39. Ronco C Bonello M Bordoni V et al. Extracorporeal therapies in non-renal disease: treatment of sepsis and the peak concentration hypothesis. Blood Purif 2004 22 164 174 Google Scholar
  • 40. Piccinni P Dan M Barbacini S et al. Early isovolaemic haemofiltration in oliguric patients with septic shock. Intensive Care Med 2006 32 80 86 Google Scholar
  • 41. Reiter K D’Intini V Bordoni V et al. High-volume hemofiltration in sepsis. Nephron 2002 92 251 258 Google Scholar
  • 42. Ronco C Tetta C Mariano F et al. Interpreting the mechanisms of continuous renal replacement therapy in sepsis: the peak concentration hypothesis. Artif Organs 2003 27 792 801 Google Scholar
  • 43. Mariano F Fonsato V Lanfranco G et al. Tailoring high-cutoff membranes and feasible application in sepsis-associated acute renal failure: in vitro studies. Nephrol Dial Transplant 2005 20 1116 1126 Google Scholar
  • 44. Ronco C Brendolan A D’Intini V Ricci Z Wratten ML Bellomo R Coupled plasma filtration adsorption: rationale, technical development and early clinical experience. Blood Purif 2003 21 409 416 Google Scholar
  • 45. Ronco C Bellomo R Acute renal failure and multiple organ dysfunction in the ICU: from renal replacement therapy (RRT) to multiple organ support therapy (MOST). Int J Artif Organs 2002 25 733 747 Google Scholar
  • 46. Cove ME MacLaren G Federspiel WJ Kellum JA Bench to bedside review: Extracorporeal carbon dioxide removal, past present and future. Crit Care 2012 21;16 5 232 Google Scholar
  • 47. Ronco C The place of early haemoperfusion with polymyxin B fibre column in the treatment of sepsis. Crit Care 2005 9 631 633 Google Scholar
  • 48. Cruz DN de Cal M Piccinni P Ronco C Polymyxin-B hemoperfusion and endotoxin removal: lessons from a review of the literature. Contrib Nephrol 2010 167 77 82 Google Scholar
  • 49. Tetta C Bellomo R Formica M et al. Use of adsorbents in ARF therapy. Contrib Nephrol 2002 137 181 8 Google Scholar
  • 50. Cantaluppi V Weber V Lauritano C et al. Protective effect of resin adsorption on septic plasma-induced tubular injury. Crit Care 2010 14 1 R4 Google Scholar
  • 51. Zhang J Peng Z Maberry D et al. Effects of hemoadsorption with a novel adsorbent on sepsis: in vivo and in vitro study. Blood Purif 2015 39 1-3 239 245 Google Scholar
  • 52. Rimmelé T Kaynar AM McLaughlin JN et al. Leukocyte capture and modulation of cell-mediated immunity during human sepsis: an ex vivo study. Crit Care 2013 26;17 2 R59 Google Scholar
  • 53. Honore PM Jacobs R Joannes-Boyau O et al. Newly designed CRRT membranes for sepsis and SIRS: a pragmatic approach for bedside intensivists summarizing the more recent advances: a systematic structured review. ASAIO J 2013 59 2 99 106 Google Scholar
  • 54. Forni LG Chawla L Ronco C Precision and improving outcomes in acute kidney injury: Personalizing the approach. J Crit Care 2017 37 244 245 Google Scholar
  • 55. Ronco C The Charta of Vicenza. Blood Purif 2015 40 1 5 Google Scholar
  • 56. Neri M Villa G Garzotto F et al. Nomenclature Standardization Initiative (NSI) alliance. Nomenclature for renal replacement therapy in acute kidney injury: basic principles. Crit Care 2016 20 1 318 Google Scholar
  • 57. Villa G Neri M Bellomo R et al. Nomenclature Standardization Initiative (NSI) Alliance. Nomenclature for renal replacement therapy and blood purification techniques in critically ill patients: practical applications. Crit Care 2016 20 1 283 Google Scholar
  • 58. Murugan R Hoste E Mehta RL et al. Acute Disease Quality Initiative (ADQI) Consensus Group. Precision Fluid Management in Continuous Renal Replacement Therapy. Blood Purif 2016 42 3 266 278 Google Scholar
  • 59. Cerdá J Baldwin I Honore PM Villa G Kellum JA Ronco C ADQI Consensus Group. Role of Technology for the Management of AKI in Critically Ill Patients: From Adoptive Technology to Precision Continuous Renal Replacement Therapy. Blood Purif 2016 42 3 248 265 Google Scholar
  • 60. Bagshaw SM Chakravarthi MR Ricci Z et al. ADQI Consensus Group. Precision Continuous Renal Replacement Therapy and Solute Control. Blood Purif 2016 42 3 238 247 Google Scholar
  • 61. Kellum JA Ronco C The 17th Acute Disease Quality Initiative International Consensus Conference: Introducing Precision Renal Replacement Therapy. Blood Purif 2016 42 3 221 223 Google Scholar
  • 62. Clark WR Garzotto F Neri M Lorenzin A Zaccaria M Ronco C Data analytics for continuous renal replacement therapy: historical limitations and recent technology advances. Int J Artif Organs 2016 39 8 399 406 Google Scholar
  • 63. Yamashita A Masaki H Kobayashi E Sukegawa T Evaluation of solute penetration across the polysulfone membrane with vitamin E coating. Hemodial Int 2015 3S S20 S25 Google Scholar
  • 64. Chen H Wu B Gong D Liu Z Fluid overload at start of continuous renal replacement therapy is associated with poorer clinical condition and outcome: a prospective observational study on the combined use of bioimpedance vector analysis and serum N-terminal pro-B-type natriuretic peptide measurement. Crit Care 2015 19 135 Google Scholar
  • 65. Rahmati S Ronco F Spittle M et al. Validation of the blood temperature monitor for extracorporeal thermal energy balance during in vitro continuous hemodialysis. Blood Purif 2001 19 2 245 250 Google Scholar
  • 66. Gura V Ronco C Nalesso F et al. A wearable hemofilter for continuous ambulatory ultrafiltration. Kidney Int 2008 73 4 497 502 Google Scholar
  • 67. Armignacco P Garzotto F Neri M Lorenzin A Ronco C Wak engineering evolution. Blood Purif 2015 39 1-3 110 104 Google Scholar
  • 68. Armignacco P Lorenzin A Neri M Nalesso F Garzotto F Ronco C Wearable devices for blood purification: principles, miniaturization, and technical challenges. Semin Dial 2015 28 2 125 130 Google Scholar
  • 69. Kim JC Garzotto F Nalesso F et al. A wearable artificial kidney: technical requirements and potential solutions. Expert Rev Med Devices 2011 8 5 567 579 Google Scholar
  • 70. Ronco C Brendolan A Bragantini L et al. Treatment of acute renal failure in newborns by Continuous Arterio-Venous Hemofiltration. Kidney Int 1986 29 4 908 915 Google Scholar
  • 71. Ronco C Evolution of Technology for Continuous Renal Replacement Therapy: Forty Years of Improvements. Contrib Nephrol 2017 189 114 123 Google Scholar
  • 72. Ronco C Ricci Z Goldstein SL Revolution in the Management of Acute Kidney Injury in Newborns. Am J Kidney Dis 2015 66 2 206 211 Google Scholar
  • 73. Ronco C Ricci Z Pediatric continuous renal replacement: 20 years later. Intensive Care Med 2015 41 6 985 93 Google Scholar
  • 74. Garzotto F Zanella M Ronco C The evolution of pediatric continuous renal replacement therapy. Nephron Clin Pract 2014 127 1-4 172 175 Google Scholar
  • 75. Ronco C Garzotto F Brendolan A et al. Continuous renal replacement therapy in neonates and small infants: development and first-in-human use of a miniaturised machine (CARPEDIEM). Lancet 2014 383 1807 1813 Google Scholar
  • 76. Ronco C Garzotto F Ricci Z CA.R.PE.DI.E.M. (Cardio-Renal Pediatric Dialysis Emergency Machine): evolution of continuous renal replacement therapies in infants. A personal journey. Pediatr Nephrol 2012 27 8 1203 1211 Google Scholar
  • 77. Kellum JA Ronco C Bellomo R Acute kidney disease and the community. Lancet 2016 14;387 10032 1974 1976 Google Scholar
  • 78. Ronco C Ricci Z De Backer D et al. Renal replacement therapy in acute kidney injury: controversy and consensus. Crit Care 2015 19 146. Google Scholar
  • 79. Bellomo R Ronco C Mehta RL et al. Acute kidney injury in the ICU: from injury to recovery: reports from the 5th Paris International Conference. Ann Intensive Care 2017 7 1 49 Google Scholar
  • 80. Ronco C Bellomo R Acute renal failure and multiple organ dysfunction in the ICU: from renal replacement therapy (RRT) to multiple organ support therapy (MOST). Int J Artif Organs 2002 25 8 733 747 Google Scholar
  • 81. Mehta R Bihorac A Selby NM et al. Acute Dialysis Quality Initiative (ADQI) Consensus Group. Establishing a continuum of acute kidney injury - tracing AKI using data source linkage and long-term follow-up: Workgroup Statements from the 15th ADQI Consensus Conference. Can J Kidney Health Dis 2016 3 13. Google Scholar
  • 82. Siew ED Basu RK Wunsch H et al. 15th ADQI Consensus Group. Optimizing administrative datasets to examine acute kidney injury in the era of big data: workgroup statement from the 15(th) ADQI Consensus Conference. Can J Kidney Health Dis 2016 3 12. Google Scholar
  • 83. Sutherland SM Chawla LS Kane-Gill SL et al. 15 ADQI Consensus Group. Utilizing electronic health records to predict acute kidney injury risk and outcomes: workgroup statements from the 15(th) ADQI Consensus Conference. Can J Kidney Health Dis 2016 3 11. Google Scholar
  • 84. Hoste EA Kashani K Gibney N et al. 15 ADQI Consensus Group. Impact of electronic-alerting of acute kidney injury: workgroup statements from the 15(th) ADQI Consensus Conference. Can J Kidney Health Dis 2016 3 10. Google Scholar
  • 85. James MT Hobson CE Darmon M et al. Acute Dialysis Quality Initiative (ADQI) Consensus Group. Applications for detection of acute kidney injury using electronic medical records and clinical information systems: workgroup statements from the 15(th) ADQI Consensus Conference. Can J Kidney Health Dis 2016 3 9 Google Scholar
  • 86. Bagshaw SM Goldstein SL Ronco C Kellum JA ADQI 15 Consensus Group. Acute kidney injury in the era of big data: the 15(th) Consensus Conference of the Acute Dialysis Quality Initiative (ADQI). Can J Kidney Health Dis 2016 3 5 Google Scholar

Authors

Affiliations

  • Department of Nephrology, Dialysis and Transplantation, St. Bortolo Hospital, Vicenza - Italy
  • International Renal Research Institute of Vicenza (IRRIV), St. Bortolo Hospital, Vicenza - Italy

Article usage statistics

The blue line displays unique views in the time frame indicated.
The yellow line displays unique downloads.
Views and downloads are counted only once per session.

No supplementary material is available for this article.